USA Lab 60L -80°C Recirculating Chiller W/ Touch Screen CLC-60/80 42L/Min
USA Lab CLC-60/80 – Recirculating Chiller (-80°C, 32L Closed Loop Reservoir)
Not recommended for continuous heat load applications such as falling film or butane recovery.
30-day warranty included.
The USA Lab CLC-60/80 is a powerful -80°C recirculating chiller featuring a 32-liter closed-loop reservoir. Built for laboratory use, it is well suited for vacuum systems, jacketed reactors, rotary evaporators, chromatography columns, solvent storage, and analytical equipment. Its larger capacity makes it especially useful for big reactors and other large vessels.
Designed for dependable performance and ease of operation, the CLC-60/80 delivers strong flow rates with a quiet-running pump, keeping up with the needs of high-demand applications. This model also comes equipped with a Programmable Logic Controller (PLC).
The PLC allows the system to switch automatically between heating and cooling modes according to programmed steps and timed cycles. For straightforward operation, it can also be set to run in static mode like a traditional chiller.
*Notice*
This unit is not compatible with 208V power input. Acceptable operating range: 219V–253V.
Not intended for continuous 24-hour use; maximum operating time is 16 hours.
In some setups, the addition of an external bypass may be required.
CLC-60/80 Features |
• Powerful chilling capacities and high BTU range. |
• High quality UL Listed compressor. |
• Touchscreen Siemens HMI display. |
• R404 refrigerant. |
• 12ft of tubing included with Tri-Clamp connections. |
• 30 Day Returns, AS-IS. |
• Versatile Siemens PLC controller with external probe. |
CLC-60/80 Specifications |
|
Expansion Tank Capacity |
32L |
Tank Dimensions |
15.75" x 7.9" x 15.75" |
Temperature Range |
RT to -80°C (-112°F) |
Power Requirements |
220-240V 60HZ Single Phase - Hardwire ONLY |
Control Type |
Siemens PLC S-700 Controller with External Probe |
Refrigeration BTUs / Wattage |
75,237BTU - 1,255BTU / 22,050-368W |
Total Power |
Nominal: 8kW / 37A | Peak: 17kW / 77A |
Plate Type Heating Area |
Heating Exchanger Surface Area 1.46m² (Quantity: 2) |
Compressor Type |
5HP Invotech (UL Listed) (Quantity: 2pcs) |
Circulating Pump Wattage |
370W (Quantity: 1) |
Fan Wattage |
280W (Quantity: 1) | 185W (Quantity: 1) |
Temperature Accuracy |
+/- 0.5°C (PT100) |
Pump Flow |
11 gpm (42L/Min) |
Lift |
91 ft (28 m) |
Pump Pressure |
40.6 psi (2.8 bar) |
Inlet / Outlet Size |
3/4" MNPT |
Refrigerant |
R404 High Stage | R23 Low Stage |
External Dimensions |
43.3" x 33.3" x 51.2" |
Weight |
514lbs |
Warranty |
None |
CLC-60/80 Packing List |
|
Item |
Quantity |
USA Lab CLC-60/80 Chiller |
1 |
Insulated Tubing |
2x 6ft (12ft Total) |
Manual |
1 |